Harmonic
SCPI Command :
[SENSe]:CORRection:CVL:HARMonic
- class HarmonicCls[source]
Harmonic commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(harmonic: float) float [source]
# SCPI: [SENSe]:CORRection:CVL:HARMonic value: float = driver.applications.k9X11Ad.sense.correction.cvl.harmonic.get(harmonic = 1.0)
Defines the harmonic order for which the conversion loss table is to be used. This setting is checked against the current mixer setting before the table can be assigned to the range. Before this command can be performed, the conversion loss table must be selected (see [SENSe:]CORRection:CVL:SELect. Is only available with option B21 (External Mixer) installed.
- param harmonic
Range: 2 to 65
- return
harmonic: No help available
- set(harmonic: float) None [source]
# SCPI: [SENSe]:CORRection:CVL:HARMonic driver.applications.k9X11Ad.sense.correction.cvl.harmonic.set(harmonic = 1.0)
Defines the harmonic order for which the conversion loss table is to be used. This setting is checked against the current mixer setting before the table can be assigned to the range. Before this command can be performed, the conversion loss table must be selected (see [SENSe:]CORRection:CVL:SELect. Is only available with option B21 (External Mixer) installed.
- param harmonic
Range: 2 to 65