IqPower
SCPI Command :
TRIGger[:SEQuence]:LEVel:IQPower
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get() float[source]
# TRIGger[:SEQuence]:LEVel:IQPower value: float = driver.applications.k91Wlan.trigger.sequence.level.iqPower.get()
Defines the magnitude the I/Q data must exceed to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. For details on the trigger source, see ‘Trigger Source’.
- Returns:
trigger_level: Range: -130 dBm to 30 dBm, Unit: DBM
- set(trigger_level: float) None[source]
# TRIGger[:SEQuence]:LEVel:IQPower driver.applications.k91Wlan.trigger.sequence.level.iqPower.set(trigger_level = 1.0)
Defines the magnitude the I/Q data must exceed to cause a trigger event. Note that any RF attenuation or preamplification is considered when the trigger level is analyzed. If defined, a reference level offset is also considered. For details on the trigger source, see ‘Trigger Source’.
- Parameters:
trigger_level – Range: -130 dBm to 30 dBm, Unit: DBM