Maximum
SCPI Command :
CALCulate<n>:LIMit<li>:BURSt:EVM:ALL:MAXimum
- Commands in total: 2Subgroups: 1Direct child commands: 1
- get(window=Window.Default, limitIx=LimitIx.Default) float[source]
# CALCulate<n>:LIMit<li>:BURSt:EVM:ALL:MAXimum value: float = driver.applications.k91Wlan.calculate.limit.burst.evm.all.maximum.get(window = repcap.Window.Default, limitIx = repcap.LimitIx.Default)
Sets or queries the average or maximum error vector magnitude limit for all carriers as determined by the default WLAN measurement. For details see ‘Modulation accuracy, flatness and tolerance parameters’.
- Parameters:
window – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)
limitIx – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Limit’)
- Returns:
limit: numeric value in dB The unit for the EVM parameters can be changed in advance using method
RsFsw.applications.k10Xlte.unit.evm.set(). Unit: DB
- set(limit: float, window=Window.Default, limitIx=LimitIx.Default) None[source]
# CALCulate<n>:LIMit<li>:BURSt:EVM:ALL:MAXimum driver.applications.k91Wlan.calculate.limit.burst.evm.all.maximum.set(limit = 1.0, window = repcap.Window.Default, limitIx = repcap.LimitIx.Default)
Sets or queries the average or maximum error vector magnitude limit for all carriers as determined by the default WLAN measurement. For details see ‘Modulation accuracy, flatness and tolerance parameters’.
- Parameters:
limit – numeric value in dB The unit for the EVM parameters can be changed in advance using method
RsFsw.applications.k10Xlte.unit.evm.set(). Unit: DBwindow – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Calculate’)
limitIx – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Limit’)
Cloning the Group
# Create a copy of the original group, that exists independently
maximum_copy = driver.applications.k91Wlan.calculate.limit.burst.evm.all.maximum.clone()
Subgroups