Band
SCPI Command :
[SENSe]:CORRection:CVL:BAND
- class BandCls[source]
Band commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get() RsFsw.enums.BandB [source]
# SCPI: [SENSe]:CORRection:CVL:BAND value: enums.BandB = driver.applications.k6Pulse.sense.correction.cvl.band.get()
Defines the waveguide band for which the conversion loss table is to be used. This setting is checked against the current mixer setting before the table can be assigned to the range. Before this command can be performed, the conversion loss table must be selected (see [SENSe:]CORRection:CVL:SELect) . Is only available with option B21 (External Mixer) installed.
- return
band: (enum or string) K | KA | Q | U | V | E | W | F | D | G | Y | J | USER Standard waveguide band or user-defined band. For a definition of the frequency range for the pre-defined bands, see Table ‘Frequency ranges for pre-defined bands’) .
- set(band: RsFsw.enums.BandB) None [source]
# SCPI: [SENSe]:CORRection:CVL:BAND driver.applications.k6Pulse.sense.correction.cvl.band.set(band = enums.BandB.D)
Defines the waveguide band for which the conversion loss table is to be used. This setting is checked against the current mixer setting before the table can be assigned to the range. Before this command can be performed, the conversion loss table must be selected (see [SENSe:]CORRection:CVL:SELect) . Is only available with option B21 (External Mixer) installed.
- param band
(enum or string) K | KA | Q | U | V | E | W | F | D | G | Y | J | USER Standard waveguide band or user-defined band. For a definition of the frequency range for the pre-defined bands, see Table ‘Frequency ranges for pre-defined bands’) .