Vswr#
SCPI Command :
- CALCulate<n>:UNCertainty:MATCh:SOURce[:VSWR]
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get ( window=Window.Default ) float
- set ( vswr: float, window=Window.Default ) None
# CALCulate<n>:UNCertainty:MATCh:SOURce[:VSWR] # get: value: float = driver.applications.k30NoiseFigure.calculate.uncertainty.match.source.vswr.get(window = repcap.Window.Default) # set: driver.applications.k30NoiseFigure.calculate.uncertainty.match.source.vswr.set(vswr = 1.0, window = repcap.Window.Default)
Defines the VSWR at the noise source output. If the noise sources during calibration and measurement are different, the command defines the uncertainty of the measurement noise source. If a smart noise source is used, the VSWR values defined in the SNS table are used.
- Parameters: