Source#
SCPI Command :
- TRIGger[:SEQuence]:SOURce
- Commands in total: 1Subgroups: 0Direct child commands: 1
- get ( ) TriggerSourceK
- set ( source: TriggerSourceK ) None
# TRIGger[:SEQuence]:SOURce # get: value: enums.TriggerSourceK = driver.applications.k17Mcgd.trigger.sequence.source.get() # set: driver.applications.k17Mcgd.trigger.sequence.source.set(source = enums.TriggerSourceK.EXT2)
Selects the trigger source. For triggering with AF, AM, AMRelative, FM, and PM trigger sources to be successful, the measurement time must cover at least 5 periods of the audio signal. For details on trigger sources, see ‘Trigger Source’. Note on external triggers: If a measurement is configured to wait for an external trigger signal in a remote control program, remote control is blocked until the trigger is received and the program can continue. Make sure that this situation is avoided in your remote control programs.
- Parameters:
source –
IMMediate: Free Run
EXTernal: Trigger signal from the Trigger Input connector.
EXT2: Trigger signal from the Trigger Input/Output connector.For FSW85 models, Trigger 2 is not available due to the second RF input connector on the front panel. The trigger signal is taken from the Trigger Input/Output connector on the rear panel.Note: Connector must be configured for Input.
EXT3: Trigger signal from the TRIGGER 3 INPUT/ OUTPUT connector.Note: Connector must be configured for Input.
RFPower: First intermediate frequency(Frequency and time domain measurements only.) Not available for input from the optional Analog Baseband interface.Not available for input from the optional Digital Baseband interface.
IFPower: Second intermediate frequencyNot available for input from the optional Digital Baseband interface.For input from the optional Analog Baseband interface, this parameter is interpreted as BBPower for compatibility reasons.
IQPower: Magnitude of sampled I/Q dataFor applications that process I/Q data, such as the I/Q Analyzer or optional applications.Not available for input from the optional Digital Baseband interface.
TIME: Time interval
BBPower: Baseband powerFor input from the optional Analog Baseband interface.For input from the optional Digital Baseband interface.
PSEN: External power sensor
AF: AF power signal
FM: FM power signal
AM: corresponds to the RF power signal
AMRelative: corresponds to the AM signal
PM: PM power signal
GP0 | GP1 | GP2 | GP3 | GP4 | GP5: For applications that process I/Q data, such as the I/Q Analyzer or optional applications, and only if the optional Digital Baseband interface is available.Defines triggering of the measurement directly via the LVDS connector. The parameter specifies which general-purpose bit (0 to 5) provides the trigger data.The assignment of the general-purpose bits used by the Digital IQ trigger to the LVDS connector pins is provided in ‘Digital I/Q’.