CmOffset#

SCPI Command :

[SENSe]:PROBe<pb>:SETup:CMOFfset
Commands in total: 1
Subgroups: 0
Direct child commands: 1
get ( probe=Probe.Default ) → float
set ( cm_offset: float, probe=Probe.Default ) → None
# [SENSe]:PROBe<pb>:SETup:CMOFfset
# get:
value: float = driver.applications.k17Mcgd.sense.probe.setup.cmOffset.get(probe = repcap.Probe.Default)
# set:
driver.applications.k17Mcgd.sense.probe.setup.cmOffset.set(cm_offset = 1.0, probe = repcap.Probe.Default)

Sets the common mode offset. The setting is only available if a differential probe in CM-mode is connected to the FSW. If the probe is disconnected, the common mode offset of the probe is reset to 0.0 V. Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode and N-mode are adapted accordingly, and vice versa. For details see .

Parameters:
  • cm_offset – Offset of the mean voltage between the positive and negative input terminal vs. ground

  • probe – optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Probe’)