State
SCPI Command :
OUTPut<up>:DIQ[:STATe]
- class StateCls[source]
State commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- get(outputConnector=OutputConnector.Default) bool [source]
# SCPI: OUTPut<up>:DIQ[:STATe] value: bool = driver.applications.k10Xlte.output.diq.state.get(outputConnector = repcap.OutputConnector.Default)
Turns continuous output of I/Q data to the optional ‘Digital Baseband’ interface on and off. Using the digital input and digital output simultaneously is not possible. If digital baseband output is active, the sample rate is restricted to 100 MHz (200 MHz if enhanced mode is possible; max. 160 MHz bandwidth) .
- param outputConnector
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Output’)
- return
state: ON | OFF | 0 | 1 OFF | 0 Switches the function off ON | 1 Switches the function on
- set(state: bool, outputConnector=OutputConnector.Default) None [source]
# SCPI: OUTPut<up>:DIQ[:STATe] driver.applications.k10Xlte.output.diq.state.set(state = False, outputConnector = repcap.OutputConnector.Default)
Turns continuous output of I/Q data to the optional ‘Digital Baseband’ interface on and off. Using the digital input and digital output simultaneously is not possible. If digital baseband output is active, the sample rate is restricted to 100 MHz (200 MHz if enhanced mode is possible; max. 160 MHz bandwidth) .
- param state
ON | OFF | 0 | 1 OFF | 0 Switches the function off ON | 1 Switches the function on
- param outputConnector
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Output’)